1011Number Key Description6Right ArrowUse to navigate right in alpha/numeric entry screens. Can be pressed simultaneously with Backlight key to incre
1213alpha/nuMerIc entryThe numeric screens will display numbers only (shown on the left). Alpha/numeric screens will display numbers acr
1415GlpThe GLP or ‘Good Laboratory Practice’ file saves detailed information about calibrations. It also includes diagnostic information
1617An example of a GLP record(Operation performed is single point % DO Calibration)*** Calibrate – DO% *** Date 02/03/09 MM/DD/YYTime
1819lanGuaGeThe Pro Plus can be configured to display all text in English, Spanish, French, German, Portuguese, Italian, Norwegian, Si
2021unIt IdUnit ID is used to identify instruments in the Data Manager software program that was included with your instrument. It is also used to id
2223SETUP DISSOLVED OXYGENPress Sensor , highlight Setup and press enter. Next, highlight DO and press enter.Enabled allows you to enable or
2425Local DO allows for localized DO% measurements. This sets the calibration value to 100% regardless of the altitude or barometric pressure. Hig
2627Auto Stable and Hold All Readings enabled, then the display will hold the readings once DO and pH have both reached their Auto Stable
2829After selecting your User ID and/or Probe ID if appropriate, highlight DO and press enter. Highlight DO % and press enter to confirm. The instrum
CONTENTSWarranty ...iIntroduction ...
3031Highlight Calibration value and press enter to manually input the sample’s dissolved oxygen value. Highlight Accept Calibration and press en
3233Highlight Calibration Value and press enter to manually enter the correct “true” barometric pressure. Next, highlight Accept Calibration, an
3435DISPLAY CONDUCTIVITYPress Sensor , highlight Display and press enter. Highlight Conductivity and press enter. Highlight Sp. Conductance (Spec
3637CALIBRATION CONDUCTIVITYThe 6051030 ISE/conductivity cable has a specialized calibration container that resembles a large test tube. This calib
3839pHpH sensors can be used on 60510-X, 6051020-X, 6051030-X, 6051010-X, and Quatro cables. If using a 605103 pH/ORP combination sensor on a
4041Sensitivity selected by User % Data Variance Threshold100 - Most Sensitive, Sensitivity bar is set to the far right0.025%75 0.39375%50 1.5%25 1.13
4243If a sensor is Enabled that isn’t connected to the instrument, the display will show an unstable false reading, ?????, or ----- next to the units.
4445Sensitivity selected by User % Data Variance Threshold100 - Most Sensitive, Sensitivity bar is set to the far right0.05%75 0.62525%50 1.275%25 1.8
4647Enabled allows you to enable or disable the ISE function and select which ISE sensor is installed on the cable. Highlight Enabled and press enter
4849CALIBRATION AMMONIUM, NITRATE, CHLORIDEThe 6051030 ISE/conductivity cable has a specialized calibration container that resembles a large test tu
Dissolved Oxygen (DO) ...21Barometer ...
5051After selecting your User ID and/or Probe ID if appropriate, highlight Ammonium, Nitrate, or Chloride to access the appropriate calibration,
5253the sky and then twist tying, zip tying, or rubber banding the bulkhead to the cable. Making a simple curve to the cable without bending or break
5455time qualifier you would like to set, press enter, and use the numeric entry screen to select the date/time you wish to view.After making your sel
5657usInG the coMMunIcatIons saddleWARNING: DO NOT connect the Communications Saddle to your PC before installing Data Manager. The Communication Sa
5859confIGure InstruMentsData Manager allows for easy and quick configuration of single or multiple instruments. Once you have uploaded a
6061GENERAL MAINTENANCE SENSOR PORTSIt is important that the entire sensor connector end be dry when installing, removing or replacing.
6263cap. If the cathode remains tarnished, contact YSI Technical Support or the Authorized dealer where you purchased the instrument.Silv
6465SENSOR MAINTENANCE CONDUCTIVITYThe openings that allow sample access to the conductivity electrodes should be cleaned regularly. The small clea
6667sensor can be regenerated by washing with alcohol and/or gently polishing with fine emery paper in a circular motion to remove any deposits or dis
6869Long-term Storage - ORPTo store, remove the sensor from the cable and seal the vacant port with the provided plug. Fill the original sh
iWAR R ANTYThe YSI Professional Plus Instrument (Pro Plus) is warranted for three (3) years from date of purchase by the end user against
7071dIssolVed oxyGenThe dissolved oxygen sensors will use Probe Current (DO uA) and Probe Slope (%/uA) as part of their GLP file records. The followi
7273Sensor Part Number Description605105**** Chloride (ISE)605106**** Nitrate (ISE)605780 Self-Stirring BOD sensor005560 Conductivity/Temperature sens
7475Accessory Part Number Description603069 Belt clip063517 Ultra clamp063507 Tripod clamp603062 Cable management kit605978 Weight, sensor/cable, 4.9
7677Harmonized Standards: •EN61326-1:2006,Electricalequipmentformeasurement, control, and laboratory use – EMC requirements – Part 1: General Req
7879The undersigned hereby declares on behalf of the named manufacturer under our sole responsibility that the listed product conforms with the Austra
Item # 605596Rev DDrawing # A605596 March 2009©2009 YSI Incorporated.
ii1INTRODUCTIONThank you for purchasing the YSI Professional Plus (Pro Plus). The Pro Plus features a waterproof (IP-67) case, backlit display and
23recommended for lab use or controlled conditions where a more rugged, field cable is not necessary. These cables include:Cable number 605107 1-me
451010 dual cablePort 1 Options Port 2 OptionspH pHORP ORPpH or pH/ORP* pH or pH/ORP*ammonium ammoniumchloride chloridenitrate nitratenone (port plug)
67Figure 2. The image on the left shows a clean, dry sensor being aligned with the bulkhead. On the right, the sensor has been pushed into the bulkh
89CONNECTING THE CABLE TO AN INSTRUMENTTo connect a cable, align the keys on the cable connector to the slots on the instrument connect