Operations ManualBODANALYSTSoftware That Makes BOD TestingEasier, More Efficient, and More Accurate
SECTION 2 SAMPLES2.1 WHAT IS A "SAMPLE?"With YSI’s BOD Analyst software you can organize your work by sample. A sample is anythingfor which
Samples Section 2YSI Incorporated BOD Analyst for Windows Software 7In this example, there are two dilutions: 150ml and 275ml. This Sample is seeded
Samples Section 2YSI Incorporated BOD Analyst for Windows Software 8Dilution Setup: You can have up to 5 dilutions for any Sample. On the General Setu
Samples Section 2YSI Incorporated BOD Analyst for Windows Software 9Quality Control: Used in the trending graph function to display Quality Control Li
Samples Section 2YSI Incorporated BOD Analyst for Windows Software 10You can find a detailed description of the abbreviation algorithm in the next sec
SECTION 3 SEEDING SAMPLES3.1 WHY SEED?Seeding is necessary when there are not enough bacteria present in the sample. A small amountof seed material t
Seeding Samples Section 3YSI Incorporated BOD Analyst for Windows Software 12If you seed directly to the bottle, the amount of seed material in each b
SECTION 4 SAMPLE GROUPS4.1 WHAT IS A "SAMPLE GROUP"?A Sample Group is a set of samples. For brevity, we often refer to Sample Groups as sim
Sample groups Section 4YSI Incorporated BOD Analyst for Windows Software 14NOTE: The dilution blank sample is automatically added when you build a bat
SECTION 5 BATCHES5.1 WHAT IS A "BATCH?"A Batch is all initial DOs that are done on any one day, along with their corresponding final DOs.Bec
Batches Section 5YSI Incorporated BOD Analyst for Windows Software 16The Open Batch Form can also be used to delete a batch. Select the batch to be d
Batches Section 5YSI Incorporated BOD Analyst for Windows Software 175.4 WORKING WITH BATCHES5.4.1 STEPS TO BUILDING A BATCHAdd the samples for the ba
Batches Section 5YSI Incorporated BOD Analyst for Windows Software 185.4.5 IMPORT BATCHThis feature allows you to setup a batch by importing a text fi
Batches Section 5YSI Incorporated BOD Analyst for Windows Software 19dilution oxygen depletion, seed correction factor and BOD are calculated automati
Batches Section 5YSI Incorporated BOD Analyst for Windows Software 20Select the sample to add from the list and the number of dilutions and the diluti
Batches Section 5YSI Incorporated BOD Analyst for Windows Software 21Notes: Used for any miscellaneous notes for the entire batch.Bottle Notes: This f
Batches Section 5YSI Incorporated BOD Analyst for Windows Software 225.14 DELETING A BOTTLETo remove a bottle from a batch, click on therow header and
SECTION 6 TAKING INITIAL DO READINGS6.1 ENTERING BOTTLE NUMBERS6.1.1 USING THE YSI 5000/5100 DO INSTRUMENTJust after selecting Read / Initials from th
Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 24then press CONFIRM.To enter "4" for the last digit,
Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 25You can also move between dilutions (bottles) without going th
iCONTENTSSECTION 1 INTRODUCTION...11.
Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 26You do not have to do the final DOs in the same order as you d
Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 27The fastest and easiest way is to set Technique: to auto. Aft
Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 28Now, we will want to measure only one of the dilutions of the
Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 29You don't have to derive initial DOs for every Sample in
SECTION 7 TAKING FINAL DO READINGS7.1 SELECTING A BATCHFrom the main bench sheet window, simply enter the sample date for the batch you would like tot
Taking Final Dos Readings Section 7YSI Incorporated BOD Analyst for Windows Software 31Your Technique for Auto Stabilize setting affects how the BOD A
Taking Final Dos Readings Section 7YSI Incorporated BOD Analyst for Windows Software 327.3 CORRECTING MISTAKESIf you have finished reading the final D
SECTION 8 REPORTSUnder the File menu, you will find the Print Daily Bench Sheet and Print Everything Reportoptions. The Print Daily Bench Sheet will p
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 348.1 REPORT DESIGNThe report designer is used to setup spreadsheet type reports.
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 35Key ActionUp Arrow Move up one rowDown Arrow Move down one rowRight Arrow Move ri
ii7.3 CORRECTING MISTAKES...32SECTION
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 36Note:To resize multiple columns at one time, select all the columns by clicking a
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 373. Release the left mouse button when the pointer is over the new location for th
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 38ABS(A1)The absolute value of the cell is returned.ISEMPTY(A1)If A1 is empty then
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 39Offset The offset from the start date of the report. For example, if the start
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 408.1.9 MONTHLY FUNCTIONSMonthly functions return summary values for a sample for a
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 41The ‘<<< Blank Report >>> selection loads a completely blank
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 428.2 EDIT MENU FUNCTIONSTo cut numbers, data values or text within the currently s
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 43The Locate Sample Formula dialog can also be opened by clicking on the Locate Sam
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 44Sample Name: Check the sample name box to have the sample name appear as a column
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 45Display [X] Decimal Places: Enter the number of digits to be displayed after the
SECTION 1 INTRODUCTIONBOD Analyst software ™ for Windows™ (BOD Analyst software) is a Windows-based PCsoftware accessory to the YSI 5000 and 5100 Diss
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 46Statistic: Sets the summary statistic to use.Decimal Places: The number of digits
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 478.4 FORMATTING CELLSThe BOD Analyst software Report Designer offers great flexibi
Reports Section 8YSI Incorporated BOD Analyst for Windows Software 488.5 OBTAINING HELP WHILE DESIGNING REPORTSThe help menu provides access to contex
SECTION 9 KEEPING YOUR DATA SAFE9.1 DATEThe BOD Analyst software depends on the correct date to do its calculations and manage thedata. The current d
Keeping Your Data Safe Section 9YSI Incorporated BOD Analyst for Windows Software 50will be created and opened. You must setup your samples and group
SECTION 10 GRAPHING10.1 INTRODUCTIONThe graphing function provides the ability to produce time series type (Data vs. Time) graphs.Up to two samples ma
Graphing Section 10YSI Incorporated BOD Analyst for Windows Software 52Sample 1 and Sample 2 are the fields in which the sample names are selected. If
Graphing Section 10YSI Incorporated BOD Analyst for Windows Software 53differences between printing in color or black and white.10.4 GETTING STATISTIC
SECTION 11 CUSTOMIZING YOUR PROGRAMThe BOD Analyst software is flexible enough to handle the wide variety of ways that people dotheir BODs. Some thing
Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 55number.The auto-stable routine starts immediately after the bot
Introduction Section 1YSI Incorporated BOD Analyst for Windows Software 2Click OK to proceed.The installation program will then display the following
Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 56Maximum acceptable value for Initial DO minus Final DO in the d
Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 57The maximum acceptable value of BOD for tests with the sample n
Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 58want the program to derive. You must measure at least one dilu
Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 59Technique: This field describes exactly how and when a stable r
Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 60Initial low temp: If the temperature of your initial DO is belo
Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 61Comm port: The program can communicate to the YSI 5000/5100 DO
Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 62Use Serial Number: Sets whether the program will assign Seria
Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 63Serial Number forRAW 10 bottle
SECTION 12 CONTACT INFORMATIONFor any questions regarding PC6000 software, or any YSI equipment, please contact YSITechnical Support at (800) 897-4151
SECTION 13 ACCESSORIESINSTRUMENTS! 5000-115 Oxygen Instrument, Digital, Lab-115V! 5100-115 Oxygen Instrument, Digital, Lab with SOUR software – 115VPR
Introduction Section 1YSI Incorporated BOD Analyst for Windows Software 3The Bench sheet is the main form of the BOD Analyst software. From the bench
1700/1725 Brannum LaneYellow Springs, Ohio 45387 USA(800) 765-4974 (937) 767-7241FAX: (937) 767-9320Website: http://www.ysi.comE-mail: [email protected]
Introduction Section 1YSI Incorporated BOD Analyst for Windows Software 4Volume of bottles (ml) 300Dilution Unit mlSeed Added to BottleReport Font Ari
Introduction Section 1YSI Incorporated BOD Analyst for Windows Software 5Coefficient of Variation Limit 0.35Apply Criteria after Final Readings OnStep