YSI Incorporated 5400 & 5500D-1, 5500D-2, 5500D-4MultiDOMonitor, Control, & Alarm InstrumentsUser Manual
YSI 5X00 182 Safety -IMPORTANT safety information. Read this chapter before installing 5X00 components and peripheral devices.is page le
YSI 5X00 199YSI 5X00 198Appendix 1 - Unpacking and InspectionInspect the outside of the shipping carton for damage. If damage is detected, co
YSI 5X00 201YSI 5X00 200Issue possible cause possible solutionDO-galvanic or polaro-graphic inaccurate valuesDO sensor wires are termi-nated i
YSI 5X00 203YSI 5X00 202Issue possible cause possible solutionSerial to ethernet device did not initialize properly. Check the event log for
YSI 5X00 205YSI 5X00 204Troubleshootingunit does not upload new code at downloaderunit is not in downloader modeSee Downloader - page 119..RS2
YSI 5X00 207YSI 5X00 206Menu MapsMenu Map Keye following conventions are used in menu maps: - Factory defaults are contained in ( ). If blan
YSI 5X00 209YSI 5X00 208Menu MapsMenu MapsCalibration*DO‐1(Enabled)DO‐2(Enabled)DO‐3(Enabled)DO‐4(Enabled)TSATmg/L*User
YSI 5X00 211YSI 5X00 210Menu MapsSystemCommunicationEthernet(NotAvailable)DHCP(Enabled)IPAddress(192.168.168.100)SubnetMask(255.255.25
YSI 5X00 213YSI 5X00 212Menu MapsRunScreen→Menus→SystemMenu→Date/TimeSystemDate/TimeSetDateFormat(MM/DD/YYYY)TDD/MM/YYYY
YSI 5X00 215YSI 5X00 214Menu MapsRunScreen→Menus→SystemMenu→GeneralAlarmSystemGeneralAlarmGeneralAlarm(Enabled)AlarmHoldOff(15S
YSI 5X00 217YSI 5X00 216Menu MapsRunScreen→Menus→SystemMenu→ServiceSystemServiceResetCleanProbesCleanProbes(Enabled)(90DaysLeft)
YSI 5X00 21YSI 5X00 20THIS IS IMPORTANTTake time to be safe when working on electrical circuits and equipment. It could save your life. Avoid
YSI 5X00 219YSI 5X00 218Menu Maps continued next pageRunScreen→Menus→SensorSetup→Aux1‐6(Analog0‐1v,0‐5v,4/20ma)SensorSetupAux1
YSI 5X00 221YSI 5X00 220Menu Maps continued next pageRunScreen→Menus→SensorSetup→Aux3‐6(Temperature)SensorSetupAux3(Temperature)
YSI 5X00 223YSI 5X00 222Menu Maps continued next pageRunScreen→Menus→SensorSetup→DO1‐4SensorSetup*DO‐1DO‐2DO‐3DO‐4DO(Enabled)Contro
YSI 5X00 225YSI 5X00 224RunScreen→Menus→SensorSetup→DO1‐4continuedfrompreviouspageSensorSetupPID/PWMcontrolonly P.I.D.Relay(
YSI 5X00 227YSI 5X00 226Menu Maps continued next pageFeedSmarttmTimer1‐4FeedTimer(Disabled)FCR(Disabled)Biomass(0oz)1to100000Feed%
YSI 5X00 229YSI 5X00 228Declaration of ConformityAppendix 4 - Declaration of ConformityThe undersigned hereby declares on behalf of the named
YSI 5X00 231YSI 5X00 230Declaration of ConformityThe undersigned hereby declares on behalf of the named manufacturer under our sole responsibi
YSI 5X00 233YSI 5X00 232Appendix 5 Parts and AccessoriesReplacement Part Part Number and DescriptionGalvanic probe/cable assembliescable lengt
YSI 5X00 235YSI 5X00 234Replacement Part Part Number and DescriptionPluggable Socket Connector for Auxil-iary Inputs, 3 pole605940UL, included
YSI 5X00 237YSI 5X00 236Appendix 7 - Alpha “A” curveAppendix 7Appendix 8 - Warrantye YSI 5X00 instrument is warranted by YSI for two years fr
YSI 5X00 23YSI 5X00 22WARNING: A switch or circuit breaker shall be included in the building where the 5X00 is installed. e disconnecting d
YSI 5X00 239YSI 5X00 238Appendix 9 - Contact UsOrdering and Technical SupportPlease visit www.ysi.com for a list of YSI’s locations worldwide
YSI 5X00 241YSI 5X00 240IndexSymbols5X00meaning & usage 10, 1230 second power up delay 77, 13730 second sensor system hold off 143AAdju
YSI 5X00 243YSI 5X00 242Indexethernet 93SMS messaging 98force serial port 102Network 98RS232 baud rate 102send test email 101Conguratio
YSI 5X00 245YSI 5X00 244force downloader 123Unit is Up-To-Date message 122Force serial port (communications menu) 102Front panelkeys 75ope
YSI 5X00 247YSI 5X00 246Service hold menu 173Submenu screen typesalpha/numeric 81numeric 81pull-down 81Systems menu 88–127adjust contrast
YSI 5X00 249YSI 5X00 248PC connection 68rubber grommets. See Bulkhead ttingsRun Screen 76run screen icons 78SSafetycautions 23warnings 2
YSI 5X00 251YSI 5X00 250biomass 165daily weight 165feed % biomass 165feed modecontinuous 162timed 163sensors 166Troubleshooting 199–205
Item # 605877Rev EDrawing # A605877August 2012Copyright©2012 YSI Incorporated.
YSI 5X00 243 Installation & Wiring -Installing & wiring 5X00 components and peripheral devices
YSI 5X00 27YSI 5X00 26Hardware InstallationCorrectly install 5X00 system components to ensure accurate data collection and reliable operation.
YSI 5X00 29YSI 5X00 28Mount the 5X00Select surface, rail, or panel mount option. Install the weather shield for outdoor installations. Surfac
YSI 5X00 31YSI 5X00 30Figure 3.42. Disconnect front panel ribbon cable by pressing the two IDC header connec-tors located on the front panel
YSI 5X00 33YSI 5X00 32Figure 3.810. Reattach lanyard by placing lanyard at proper location of display board panel. Make sure lanyard is full
YSI 5X00 35YSI 5X00 34For accurate DO measurements, ensure that the temperature reading for each DO channel is representative of the temperatu
YSI 5X00 37YSI 5X00 36• Since the ProODO utilizes optical luminescent technology, continuous sample movement or stirring is not required. (Pl
YSI 5X00 3YSI 5X00 2Item # 605877Rev EDrawing # A605877August 2012Copyright©2012 YSI Incorporated.All trademarks herein areregistered to YSI I
YSI 5X00 39YSI 5X00 38Figure 3.12YSI 626250 ODO (Optical Dissolved Oxygen) sensore YSI 626250 probe assembly includes an ODO sensor and senso
YSI 5X00 41YSI 5X00 40Notes: - I/O board and control board (including DO daughter board) are attached to a metal plate and are referred to as
YSI 5X00 43YSI 5X00 42Accessing the I/O Boarde I/O board is accessible when the 5X00 front panel is removed. e letters on the metal plate, w
YSI 5X00 45YSI 5X00 44Control Boarde control board is mounted underneath the I/O Board. e serial to ethernet device and lithium battery are
YSI 5X00 47YSI 5X00 467. Terminate other end of the earth ground wire to a cold water pipe, steel rod, copper pipe, or earth ground.8. Compl
YSI 5X00 49YSI 5X00 48Wire DC Power and Backup Powere 5X00 may be powered by most regulated 9.0-16.5 VDC source that provides 800mA of curren
YSI 5X00 51YSI 5X00 50Step 5 Wire Sensor(s)Up to four DO sensors and four temperature sensors can be wired to the 5X00. All sensors are wired
YSI 5X00 53YSI 5X00 52Do not connect or disconnect IDC ribbon cablewhen 5400 is powered. Serious damage can occur.ground DOs
YSI 5X00 55YSI 5X00 545. Wire temperature sensors. Temperature sensors are wired as auxiliary inputs. See Temperature Sensor(s) - page 54.6.
YSI 5X00 57YSI 5X00 56Step 6 Wire Relay outputse 5X00 has eight (8) on board non-fused relays rated at 240VAC@5AMPs (110VAC@10AMPs) of switch
YSI 5X00 5YSI 5X00 4Table of ContentsStep 4 Wire Power ...46Wire AC Powe
YSI 5X00 59YSI 5X00 58Wiring a Contactore eight onboard relays are rated at 240VAC@5AMPs (110VAC@10AMPs) switch-ing capacity. Contactors sh
YSI 5X00 61YSI 5X00 60See Digital Input example - gure 3.24 and Analog Input example - gure 3.25.CAUTION: e auxiliary inputs are not isola
YSI 5X00 63YSI 5X00 62Step 8 Wire NetworkMultiple 5X00 and 5200A instruments can be networked using the RS485 network ports. A node network c
YSI 5X00 65YSI 5X00 64Note: - It is essential that wires be terminated cleanly. Nicks (stray wires, not cleanly terminated) can result in unr
YSI 5X00 67YSI 5X00 66masterTesting network configurationsTesting network configurationsslavemasterslaveslave slave-represents wired network t
YSI 5X00 69YSI 5X00 68Notes: - Firmware can only be uploaded to an instrument using a local RS232 connec-tion. See Downloader - page 119. - O
YSI 5X00 71YSI 5X00 70Ethernet (TCP/IP) Module Communicatione optional serial to ethernet device must be installed on the 5X00 control board
YSI 5X00 724 Conguring 5X00 Systems -Front panel keys-Run Screen-Navigating menus-Menu structure-Conguring 5X00 systemsis page le
YSI 5X00 75YSI 5X00 745X00 Conguration and Monitoringis chapter provides information on the 5X00’s front panel, navigating menus, and menu f
YSI 5X00 77YSI 5X00 76Sokey label FunctionExit menu return to Run Screen saving menu conguration changes.Exit menu (to previous menu) withou
YSI 5X00 7YSI 5X00 6Range High Timed and Normal Control ...151Directions to Congure Temperature System ...
YSI 5X00 79YSI 5X00 78Run Screen IconsIcons identify control, alarm, timer, power, service, and communication informa-tion. Icons along the t
YSI 5X00 81YSI 5X00 80Menus5X00 rmware is structured using menus that enable you to easily congure and view system status, messages, and log
YSI 5X00 83YSI 5X00 82Main MenuPress at the Run Screen to enter the Main Menu. All system conguration is done through this menu. e four M
YSI 5X00 85YSI 5X00 84Figure 4.7Notes: - A sensor must be enabled in order to calibrate. - For YSI 5422 and YSI 626250 cable/probe assemblies,
YSI 5X00 87YSI 5X00 8613. Enter DO calibration menu Menus → Calibration→ DO→Sat14. Allow approximately 10 minutes for the sensor to settl
YSI 5X00 89YSI 5X00 88Note: - Elevation is not required for mg/L Calibration.System MenuMenus → Systeme 5400 System Menu contains 17 submenus
YSI 5X00 91YSI 5X00 901. Enter BackLight Timer menu Menus → System →BackLight Timer2. Use▲and ▼ to scroll and highlight submenus.3. Press
YSI 5X00 93YSI 5X00 92messages. Verify with the service provider and/or network administrator to conrm that the 5X00 communication congurat
YSI 5X00 95YSI 5X00 94 1. Install serial to ethernet device - page 65.2. e Ethernet icon displays at the Run Screen when the serial to et
YSI 5X00 97YSI 5X00 96Subnet MaskMenus → System → Communications → Ethernet → Subnet MaskA Subnet mask address can only be congured when DHCP
YSI 5X00 81 Introduction -5400/5500 Product Description-User Manual Guide-Specications is page le intentionally blank.
YSI 5X00 99YSI 5X00 98potentially delay any AquaManager alarm emails from being sent. - To disable the serial to ethernet device from sending
YSI 5X00 101YSI 5X00 1009. Press to save conguration and return to Run Screen. 10. Each slave should be tested one by one. Insert network
YSI 5X00 103YSI 5X00 102Email Retry DelayMenus → System → Communications → Email Retry DelayCongure the Email Retry Delay for the time interv
YSI 5X00 105YSI 5X00 104 5400 Congure Aux 5 menu 5500 Congure Aux 5 menu(no Temp Source menu) (Temp Source menu will be available w
YSI 5X00 107YSI 5X00 106Notes: - It is normal to hear an on-board relay “click” when switching to or from the 4-20ma aux input type. - Analog
YSI 5X00 109YSI 5X00 108gure 4.7Notes: - Values displayed at the run screen are based on display format resolution conguration. Example - i
YSI 5X00 111YSI 5X00 110 Data LoggingMenus → System →Data Logginge 5X00 saves information in calibration, sensor, and event logs. Press
YSI 5X00 113YSI 5X00 112 In the above sensor log screen shot Aux 3 is assigned as temperature for DO 1. DO 4 and Aux 1, 2, 4, 5, & 6 are d
YSI 5X00 115YSI 5X00 114Date/TimeMenus → System → Date/TimeIt is important to congure the correct date and time in the 5X00. Four 10-event d
YSI 5X00 117YSI 5X00 1161. Enter Display Settings menu Menus → System → Display Settings →Display Format2. Use▲and ▼ to scroll and highlig
YSI 5X00 11YSI 5X00 10Introductionank you for purchasing the YSI 5X00. roughout this manual YSI “5X00” is used globally to refer to YSI 540
YSI 5X00 119YSI 5X00 1185. Aer entering all K values press . is initiates a validation of the entered values. (Do NOT press sokey.
YSI 5X00 121YSI 5X00 1205. Congure PC com port number (1-8) in Flash Programmer.Notes: - Use only a Tripp-Lite Keyspan USB-Serial Adapter,
YSI 5X00 123YSI 5X00 122Notes: - During upload, 5X00 systems are placed into Service Hold which de-energizes any active systems including ener
YSI 5X00 125YSI 5X00 124General AlarmMenus → System → General Alarme General Alarm triggers the buzzer, general alarm relay, general alarm ic
YSI 5X00 127YSI 5X00 126Names can contain up to 20 characters. Use AquaManager to easily congure names using the PC keyboard. See Aqua-Mana
YSI 5X00 129YSI 5X00 128 - displays at Run Screen when password is enabled.Power Fail MenuMenus → System → Power FailIn the event of AC powe
YSI 5X00 131YSI 5X00 130Reset Systems to factory defaults, and test Relay(s). Reset Clean ProbesMenus → System → Service→ Reset Clean ProbesEn
YSI 5X00 133YSI 5X00 1324. Press to save conguration and return to Run Screen.Notes: - Display contrast, backlight timer, and buzzer can on
YSI 5X00 135YSI 5X00 134Relay TestMenus → System → Service→ Relay TestUse the Relay Test menu to verify peripheral devices are wired correctly
YSI 5X00 137YSI 5X00 136AquaManager and the System MenuAquaManager users can congure most 5X00 system menus from a PC. Aer polling the 5X00,
YSI 5X00 13YSI 5X00 12Figure 1.1 Preview of User Manualis manual includes information on installing, conguring, operating, and main-tai
YSI 5X00 139YSI 5X00 138 - One relay can be assigned to multiple systems. It is essential that the relay conguration be veried in order to
YSI 5X00 141YSI 5X00 140Range #1LowAlarmONHighAlarmONHighRange #1ONOptimalSetpointValueLowerParameterValuesHighAlarmSetpointLowAlarmSetpointHi
YSI 5X00 143YSI 5X00 142Example - entering “63” for DO set point results in validation error because the valid range for the DO sensor system
YSI 5X00 145YSI 5X00 144PID/PDW TuningEach and every water system responds very dierently to peripheral control de-vices due to many factors.
YSI 5X00 147YSI 5X00 146PID/PWM Alarm SystemAlarm values set the acceptable control range. Enable and congure system(s) alarm to activate al
YSI 5X00 149YSI 5X00 148Aux Relay DigitalMenus → Sensor Setup→Aux 1-6When enabled, Aux digital relay energizes when congured control/alarm on
YSI 5X00 151YSI 5X00 150Directions to Congure Aux (non temperature analog)1. Wire Aux analog devices - page 59.2. Select “0-1v”, “0-5v”, or
YSI 5X00 153YSI 5X00 152Directions to Congure Temperature System 1. Wire temperature sensor to 5X00 I/O Board (for non ODO sensor temperatu
YSI 5X00 155YSI 5X00 154 - Aux temperature and ODO temperature values are not accessible at the 5X00. Temperature readings can only be obtain
YSI 5X00 157YSI 5X00 1561. Install and wire DO probe(s) and membrane and/or sensor cap - page 50.2. Congure DO system temperature source :
YSI 5X00 15YSI 5X00 14Input System SpecicationsAux Systemsere are 6 Aux input systems that allow digital or analog (0-1.0VDC, 0-5.0VDC, or 4
YSI 5X00 159YSI 5X00 158AquaManager and the Sensor Setup MenuAll Sensor Setup menus can be congured using AquaManager. Aer polling the 5X00
YSI 5X00 161YSI 5X00 1606. Press to save conguration and return to Run Screen.Notes: - Timer system must be enabled in order to energize
YSI 5X00 163YSI 5X00 162 - feed duration - length of time feeder relay is energized; - feed interval - length of time between feedings. Option
YSI 5X00 165YSI 5X00 164Conguring the 5X00 - e Feed timer will not operate properly if the start date is set prior to Janu-ary 1, 2007. Dail
YSI 5X00 167YSI 5X00 166SensorsMenus →Timers→Feed Smart Timer→Feeder→Sensors →Disabled/Enabled →DO Low/High/
YSI 5X00 169YSI 5X00 1686. Press for feed timer relay to energize for congured run test time. Countdown of relay “on” time is displayed.7
YSI 5X00 171YSI 5X00 170 - e Run Screen must be displayed during normal operation. If the 5X00 is not at the run screen during AquaManager u
YSI 5X00 173YSI 5X00 172Note: - User congured sensor and relay names do not appear in log menus.Calibration Logs Calibration logs store senso
YSI 5X00 174systems are reset causing relays to energize/de-energize based on the system value/condition. Timer systems are not reset. To reset
YSI 5X00 177YSI 5X00 176Installing AquaManagerAquaManager provides data logging, graphing, autopolling, mapping, system conguration control,
YSI 5X00 17YSI 5X00 16Component and System (non input) SpecicationsHardwareSystem System Feature Specication Display/LEDs backlit graphi
YSI 5X00 179YSI 5X00 178See AquaManager Online Help for detailed application conguration information. Before the node can be accessed using
YSI 5X00 181YSI 5X00 180 - access AquaManager Help - check for AquaManager soware updates - AquaManager about informationNode (Graph) WindowD
YSI 5X00 183YSI 5X00 182Database ManagementFor optimal operation and safeguarding, compact and back up the AquaManager database. Uncompacted
YSI 5X00 185YSI 5X00 184MappingCreate a network map to display current conditions* for 5X00 nodes at PC run-ning AquaManager. Map icons are c
YSI 5X00 186is page le intentionally blank.6 Maintenance -Sensor Calibration Checks-Sensor Maintenance-Sensor Storage-Charging DC batteri
YSI 5X00 189YSI 5X00 188MaintenanceIt is important to remember that the quality of data is dependent on three factors directly related to the
YSI 5X00 191YSI 5X00 1904. Remove the probe from the sample stream.5. Remove the probe guard from the 5422 and 626250 probes. 6. Manually r
YSI 5X00 193YSI 5X00 192Before reinstalling the sensor cap, make sure that the cavity is completely drybefore proceeding with the installation
YSI 5X00 195YSI 5X00 194Figure 6.24. Remove and dispose of the membrane and its O-Ring as shown above. 5. Using a toothbrush, dish washing p
YSI 5X00 196Long Term Storage 626250Long term storage is dened as a period of time > 30 days. e 626250 sensor MUST BE STORED IN A MOIST ENVIR