ba76033e01 01/2012IQ SENSOR NETMIQ/C6Operating ManualPower!OKIQ SENSOR NET six-channel current output module
Installation MIQ/C63 - 2ba76033e01 01/20123.3 Electrical connections: General instructionsCable glands All electric cables are fed from below via prep
MIQ/C6 Installation3 - 3ba76033e01 01/2012General installationinstructionsObserve the following points when attaching connecting wires to the terminal
Installation MIQ/C63 - 4ba76033e01 01/2012Connecting lines to theterminal stripFig. 3-1 Terminal strip with the current connections1 Open the module.2
MIQ/C6 Settings4 - 1ba76033e01 01/20124 SettingsThe MIQ/C6 module has six current outputs that supply a current that depends on the measured value.On
Settings MIQ/C64 - 2ba76033e01 01/20124.1 Entering / editing the name of an output For easier identification of the outputs, an individual name can be
MIQ/C6 Settings4 - 3ba76033e01 01/20124.2 Linking the output with a sensor Fig. 4-2 150 - Settings of outputs and links: Link with...NoteOutputs that
Settings MIQ/C64 - 4ba76033e01 01/20124.3 Deleting a link with an outputIf a link from a current output with a sensor is no longer required, the link
MIQ/C6 Settings4 - 5ba76033e01 01/20124.4 Setting current outputsFig. 4-4 150 - Settings of outputs and links1 Call up the measured value display with
Settings MIQ/C64 - 6ba76033e01 01/20124.4.1 RecorderFunction In the Recorder application, the measured values of the linked sensor are set up as curre
MIQ/C6 Settings4 - 7ba76033e01 01/2012Behavior at error Fixed current valueIn the case of system and sensor errors, the current output supplies the sp
MIQ/C62ba76033e01 01/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs, OH 45387
Settings MIQ/C64 - 8ba76033e01 01/20124.4.2 PID controllerFunction With the PID controller function, you can use an output as a controller output. The
MIQ/C6 Settings4 - 9ba76033e01 01/2012By activating or deactivating the Integral (ti) and Derivative (td) controller part, the following controller ty
Settings MIQ/C64 - 10ba76033e01 01/2012Application example Feedback control of the D. O. concentration Sensor: TriOxmatic 700 IQ (measuring range:
MIQ/C6 Settings4 - 11ba76033e01 01/2012Within the proportional band, a concentration increase by 1 mg/l causes a decrease of the current output by 1 m
Settings MIQ/C64 - 12ba76033e01 01/2012Io 0 ... 20 mA Current value on the output if the measured value equals the Nominal valueti 0 ... 9999 s Hold-b
MIQ/C6 Settings4 - 13ba76033e01 01/20124.4.3 Fixed current valueFunction With the Fixed current value function, you can test the operativeness of the
Settings MIQ/C64 - 14ba76033e01 01/20124.5 Checking the status of the outputsThis function offers a simple overview of the states of all outputs of a
MIQ/C6 Settings4 - 15ba76033e01 01/20124.6 Behavior of linked outputs4.6.1 Behavior in case of errorFor linked current outputs, you can determine the
Settings MIQ/C64 - 16ba76033e01 01/20124.6.2 Behavior in the non-operative conditionIn the non-active condition the output current is 0 mA.An output i
MIQ/C6 Settings4 - 17ba76033e01 01/20124.7 Maintenance condition of sensorsThe maintenance condition of sensors serves to calibrate, clean, service an
MIQ/C6 List of contents0 - 1ba76033e01 01/2012MIQ/C6 - List of contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Settings MIQ/C64 - 18ba76033e01 01/20124.7.1 Switching on the maintenance condition4.7.2 Switching off the maintenance condition1 Call up the measured
MIQ/C6 Maintenance and cleaning5 - 1ba76033e01 01/20125 Maintenance and cleaning5.1 MaintenanceThe MIQ/C6 requires no special maintenance. The general
Maintenance and cleaning MIQ/C65 - 2ba76033e01 01/2012
MIQ/C6 Technical data6 - 1ba76033e01 01/20126 Technical dataNoteGeneral technical data on MIQ modules are given in the TECHNICAL DATA chapter of the I
Technical data MIQ/C66 - 2ba76033e01 01/2012Current outputsOutput Physically separated from the IQ SENSOR NETOutput current Can be switched between 0
MIQ/C6 Contact Information7 - 1ba76033e01 01/20127 Contact Information7.1 Ordering & Technical SupportWhen placing an order please have the follow
Contact Information MIQ/C67 - 2ba76033e01 01/2012
MIQ/C6 Lists8 - 1ba76033e01 01/20128Lists8.1 Explanation of the messagesIn this chapter you will find a list with all the message codes and correspond
Lists MIQ/C68 - 2ba76033e01 01/2012
List of contents MIQ/C60 - 2ba76033e01 01/20127.2 Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-18 Lists . . . . . .
1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com
MIQ/C6 Overview1 - 1ba76033e01 01/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NEToperating manualFig. 1-1 Str
Overview MIQ/C61 - 2ba76033e01 01/20121.2 Features of the MIQ/C6General characteristics The MIQ/C6 output module has six current outputs. They can be
MIQ/C6 Safety instructions2 - 1ba76033e01 01/20122 Safety instructionsThis operating manual contains special instructions that must be followed during
Safety instructions MIQ/C62 - 2ba76033e01 01/20122.1 Authorized useThe authorized use of the MIQ/C6 consists of providing current outputs in the IQ SE
MIQ/C6 Installation3 - 1ba76033e01 01/20123 Installation3.1 Scope of deliveryThe scope of delivery of the MIQ/C6 is listed in the INSTALLATION chapter