Operations Manual YSI ADV6600 Environmental Monitoring System

Seite 4

ADV6600 Y S I Environmental Page 91 Section 8. ADVantage 6600 Software – Post-Processing of Data 8-1. Opening a Saved File After a file

Seite 5

Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 92 format of the default template. Note, however, that each time you open a

Seite 6

Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 93 7. The software will now ask you to choose a name for the new file and s

Seite 7

Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 94 1 Left Click and Hold Moves marker to change data point being displayed

Seite 8

Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 95 8-4. Toolbars ADVantage 6600 software utilizes toolbars for some of the

Seite 9

Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 96 8-4.4. Data Collection Toolbar Utilities Future feature – not present

Seite 10

Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 97 Report Opens the Report window for viewing of File Summary, User Settings

Seite 11 - 1-2. About YSI, Inc

Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 98 Tabular Data Displays or hide the Tabular Data at the bottom of the scre

Seite 12 - 1-3. How to Use This Manual

Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 99 Hardware Configuration Displays hardware specifications of the instrumen

Seite 13 - 1-5. Safety Considerations

ADV6600 Y S I Environmental Page 1 Section 1. Introduction to the ADV6600 1-1. About the ADV6600 The ADV6600 is a fully integrated syste

Seite 14

Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 100 8-5.6. Window ADVantage 6600 software allows for multiple data files t

Seite 15 - Environmental Page 5

ADV6600 Y S I Environmental Page 101 Section 9. Principles of Operation 9-1. Acoustic Doppler Velocimeter (ADV) The ADV6600 measures the

Seite 16 - Environmental Page 6

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 102 (sediment, small organisms, bubbles, etc.). Some portion of the reflec

Seite 17 - Environmental Page 7

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 103 9-2.2. XYZ Using the XYZ coordinate system, velocity measurements are

Seite 18 - Environmental Page 8

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 104 Where: dBar = measured pressure (decibar) PressOffset = offset calibra

Seite 19 - 2-2.2. Bulkhead Diagram

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 105 mS/cm, which have been prepared in accordance with recommendation 56-19

Seite 20

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 106 9-8. TDS Procedure from Standard Methods for the Examination of Wat

Seite 21 - Environmental Page 11

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 107 detectors of DO are highly flow-dependent and therefore require externa

Seite 22 - 2-3.2. Preparing the Cable

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 108 which is necessary to obtain relatively flow independent measurements.

Seite 23 - Sonde to Computer

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 109 9-10. pH Procedure from Standard Methods for the Examination of Water

Seite 24

Section 1. Introduction to the ADV6600 ADV6600 Y S I Environmental Page 2 environmental enforcement personnel, officials of water utilities, an

Seite 25 - Sonde Firmware Set-up

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 110 intercept at Tm (temperature of measurement). Fortunately, the Nernst

Seite 26 - Environmental Page 16

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 111 WARNING: Nitrate sensors can only be used at depths of less than 50 f

Seite 27 - Environmental Page 17

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 112 indicates the presence of only small quantities of nitrate, it is unlik

Seite 28 - Environmental Page 18

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 113 temperature at least 10°C different from the first two calibration poin

Seite 29 - Environmental Page 19

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 114 whichever is larger) refers to sampling applications where only minimal

Seite 30 - SONDE MENU

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 115 and the quality of the water. For all monitoring studies using ion sel

Seite 31 - 3-6.4. Setting Up Reports

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 116 (2) The absolute values of the readings are significantly closer to th

Seite 32 - Environmental Page 22

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 117 particles in the optical path at the instant of measurement. For examp

Seite 33 - Environmental Page 23

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 118 The limitations of the in vivo method are outlined below and should be

Seite 34 - Environmental Page 24

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 119 The unique YSI chlorophyll system available as an option for use with t

Seite 35 - Environmental Page 25

Section 1. Introduction to the ADV6600 ADV6600 Y S I Environmental Page 3 If you have any questions about this product or its application, plea

Seite 36 - Environmental Page 26

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 120 From our studies, it appears that entry of a value of 1 to 2 % per degr

Seite 37 - Environmental Page 27

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 121 As noted above, the measurement of chlorophyll from in vivo fluorescenc

Seite 38 - Environmental Page 28

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 122 provide an accuracy specification for chlorophyll measurement made with

Seite 39 - Environmental Page 29

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 123 As for chlorophyll, rhodamine WT fluoresces, that is, when irradiated

Seite 40 - Environmental Page 30

Section 9. Principles of Operation ADV6600 Y S I Environmental Page 124 9-17.1. Calibration and Effect of Temperature The sonde software offe

Seite 41 - Environmental Page 31

ADV6600 Y S I Environmental Page 125 Section 10. Care, Maintenance, and Storage 10-1. Protection from Biological Fouling The ADV6600 can

Seite 42 - Environmental Page 32

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 126 Within the United States, it is recommended to use an anti-foul

Seite 43 - Environmental Page 33

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 127 have completed painting the conductivity cell of your probe, it

Seite 44

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 128 10-2. Sonde Care and Maintenance When caring for your ADV6600

Seite 45 - 4-1. Beam Check Basics

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 129 2. Draw the o-ring through the grease while pressing the finge

Seite 47 - Environmental Page 37

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 130 10-3. Probe Care and Maintenance 10-3.1. ADV Biological gro

Seite 48 - Environmental Page 38

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 131 To resurface the probe: 1. First dry the probe tip completely

Seite 49 - Environmental Page 39

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 132 If good pH and/or ORP response is still not restored by the ab

Seite 50 - Environmental Page 40

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 133 water under examination. A replacement wiper is supplied with

Seite 51 - Environmental Page 41

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 134 10-5.3. Temperature Probe No special precautions are required

Seite 52 - Environmental Page 42

Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 135 The key to pH probe storage, short or long-term, is to make cer

Seite 54 - Environmental Page 44

ADV6600 Y S I Environmental Page 137 Section 11. Troubleshooting This section contains troubleshooting information that will be helpful to

Seite 55 - Environmental Page 45

Section 11. Troubleshooting ADV6600 Y S I Environmental Page 138 11-2. Communication Problems 11-2.1. Cannot Communicate With ADV6600 In th

Seite 56 - Environmental Page 46

Section 11. Troubleshooting ADV6600 Y S I Environmental Page 139 Low scattering strength If there is insufficient scattering material (e.g.,

Seite 57 - 4-5.3. Calibration Tips

ADV6600 Y S I Environmental Page 5 Section 2. Preparing the System for Field Studies Before using your ADV6600 in field studies to correla

Seite 58 - Environmental Page 48

Section 11. Troubleshooting ADV6600 Y S I Environmental Page 140 Malfunctioning receiver(s) The signal from each receiver should have about t

Seite 59 - Environmental Page 49

Section 11. Troubleshooting ADV6600 Y S I Environmental Page 141 Excessive Noise If the electronics noise level is more than 10 counts above i

Seite 60 - 4-6.2. Conductivity

Section 11. Troubleshooting ADV6600 Y S I Environmental Page 142 • Be certain the sampling volume of the instrument is sufficiently spaced fro

Seite 61 - Environmental Page 51

Section 11. Troubleshooting ADV6600 Y S I Environmental Page 143 SYMPTOM POSSIBLE CAUSE ACTION REFProbe not properly calibrated Follow DO

Seite 62 - 4-6.5. pH

Section 11. Troubleshooting ADV6600 Y S I Environmental Page 144 SYMPTOM POSSIBLE CAUSE ACTION REFConductivity improperly calibrated. Fol

Seite 63 - 4-6.6. ORP

ADV6600 Y S I Environmental Page 145 Section 12. Warranty and Service Information Thank you for purchasing a YSI ADV6600. This instrument

Seite 64 - 4-6.7. Ammonium

Section 12. Warranty and Service Information ADV6600 Y S I Environmental Page 146 EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WA

Seite 65 - Environmental Page 55

Section 12. Warranty and Service Information ADV6600 Y S I Environmental Page 147 ) If exposure has occurred, the customer must certify that d

Seite 67 - 4-6.9. Chloride

ADV6600 Y S I Environmental Page 149 Product Return Form Customer Information Organization _______________________ Contact Name ______________

Seite 68 - Environmental Page 58

Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 6 • Hold the probe in a vertical position and apply a few drops of electr

Seite 69 - 4-6.10. Turbidity

ADV6600 Y S I Environmental Page 150 Product Return Form _________________________________ City, State, Zip _____________________ _____________

Seite 70 - Environmental Page 60

ADV6600 Y S I Environmental Page 151 Section 13. Additional Support YSI Environmental offers a wide range of customer assistance and techni

Seite 72 - Calibration

ADV6600 Y S I Environmental Page 153 Appendix A. Accessories and Calibration Standards A-1. Probes and Probe Replacement Parts YSI Model Nu

Seite 73 - Environmental Page 63

Appendix A. Accessories and Calibration Standards ADV6600 Y S I Environmental Page 154 A-2. Optional Accessories and Replacement Parts YSI Mo

Seite 74 - 4-6.12. Rhodamine WT

Appendix A. Accessories and Calibration Standards ADV6600 Y S I Environmental Page 155 YSI Model Number Description 6656 ADVantage 6600 Windo

Seite 75 - Environmental Page 65

Appendix A. Accessories and Calibration Standards ADV6600 Y S I Environmental Page 156 A-4. Reagents YSI Model Number Description 3161 Cond

Seite 76 - Command

ADV6600 Y S I Environmental Page 157 Appendix B. Specifications Sensor Range Resolution Accuracy Velocity ±0.001 to 6 m/s (±0.003 to 19 ft/s

Seite 77 - 5-1. Mounting Methods

Appendix B. Specifications ADV6600 Y S I Environmental Page 158 Sensor Range Resolution Accuracy Rhodamine WT 0 to 200 µg/L as true dye; 1,00

Seite 78 - Environmental Page 68

ADV6600 Y S I Environmental Page 159 Appendix C. Required Notice The Federal Communications Commission defines this product as a computing de

Seite 79 - Environmental Page 69

Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 7 • Remove the old DO membrane and clean the probe tip with water and le

Seite 81 - Environmental Page 71

ADV6600 Y S I Environmental Page 161 Appendix D. Frequently Asked Questions D-1. System Description and General Questions Can you give m

Seite 82 - Environmental Page 72

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 162 Is it possible to get the ADV6600 in an “EDS” configuration where

Seite 83 - 5-2. Mounting Cautions

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 163 Can I use my existing YSI and/or SonTek field cables when communica

Seite 84 - Environmental Page 74

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 164 D-2. What does the system measure? What is the difference betwee

Seite 85 - 5-2.2. Flow Interference

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 165 • The sensor must be totally submerged to successfully measure th

Seite 86

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 166 D-3. Installing the System How should I deploy my ADV6600 for lo

Seite 87 - Environmental Page 77

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 167 A few examples of good deployment conditions are: 1. Mounting th

Seite 88 - Environmental Page 78

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 168 Can I deploy my ADV6600 by attaching it to a metal pole or mount, o

Seite 89 - Environmental Page 79

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 169 2. Spot sampling/profiling of these current and water quality para

Seite 90 - Environmental Page 80

Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 8 • Roll the o-ring over the end of the probe, being careful not to touc

Seite 91 - Environmental Page 81

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 170 Yes. The standard compass that is present in each ADV6600 will auto

Seite 92 - 6-4. SDI-12 Data Collection

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 171 If I download data to my PC periodically in the field and restart

Seite 93 - Environmental Page 83

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 172 A Microsoft Windows operating system is required, and it must be Wi

Seite 94 - Environmental Page 84

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 173 No. The released version of ADVantage 6600 has no data merging capa

Seite 95 - Environmental Page 85

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 174 No. The ADV sensor does not require calibration. However, if the ar

Seite 96

Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 175 YSI offers a wide range of customer assistance and technical suppor

Seite 98 - Environmental Page 88

ADV6600 Y S I Environmental Page 177 Appendix E. Chlorophyll Measurements This section is designed to help the user attain the maximum poss

Seite 99 - Environmental Page 89

Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 178 It is important for the user to remember that, because of the limit

Seite 100

Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 179 time of field measurement for the chlorophyll values determined by th

Seite 101 - Processing of Data

Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 9 2-2.2. Bulkhead Diagram  ADV Probe = Non-removable  Pressure Sensor

Seite 102 - Environmental Page 92

Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 180 Are there precautions that I should take when acquiring samples for

Seite 103 - Environmental Page 93

Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 181 analytical procedure has begun. Some users prefer to filter the samp

Seite 104 - Environmental Page 94

Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 182 If you consistently observe periods in your monitoring study where th

Seite 105 - 8-4. Toolbars

Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 183 Note the significant improvement in the overall noise level of the

Seite 106 - 8-5. Menu Features

Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 184 temperature at the site is 10 C, the observed chlorophyll readings wi

Seite 107 - 8-5.2. Edit

ADV6600 Y S I Environmental Page 185 Appendix F. Percent Air Saturation The term “Percent Air Saturation” (abbreviated in many applications

Seite 108 - 8-5.4. System

Appendix F. Percent Air Saturation ADV6600 Y S I Environmental Page 186 F-2. “DOsat % Local” Convention The second convention has been used

Seite 109 - 8-5.5. Tools

Appendix F. Percent Air Saturation ADV6600 Y S I Environmental Page 187 F-3. Effects of DO mg/L Calibration When a user performs a mg/L cali

Seite 110 - 8-5.7. Help

Appendix F. Percent Air Saturation ADV6600 Y S I Environmental Page 188 11. Select Change System Settings. A Quick Setup dialog box will app

Seite 111 - Environmental Page 101

ADV6600 Y S I Environmental Page 189 Appendix G. Quick Start Deployment Guide 1-Installing the ADVantage 6600 Software A CDROM containing the

Seite 113 - 9-4. Pressure

Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 10 2-2.3. O-Ring Lubrication Apply a very thin coat of o-ring lubricant,

Seite 114 - 9-6. Conductivity

Appendix G. Quick Start Deployment Guide ADV6600 Y S I Environmental Page 190 To reduce the amount of space taken up by files on the recorder

Seite 116 - 9-9. Dissolved Oxygen

ADV6600 Y S I Environmental Page 192 YSI Environmental 1700/1725 Brannum Lane Yellow Springs, OH 45387 USA 937.767.7241 937.767.935

Seite 117 - 9-9.1. Method of Operation

Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 11 Any ISE probe can be installed in either of the two ports labeled “3”

Seite 118 - 9-9.3. Flow Dependence

Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 12 2-3.2. Preparing the Cable The cable purchased with your ADV6600 has

Seite 119 - 9-10. pH

Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 13 2-3.4. Attaching the Cable Attach your ADV6600 to a computer for ini

Seite 121 - Environmental Page 111

ADV6600 Y S I Environmental Page 15 Section 3. Configuring Your ADV6600 – Installation of ADVantage 6600 Software and Sonde Firmware Set-up

Seite 122 - 9-13. Ammonium and Ammonia

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 16 3-4. Connecting to the System When choosing the “Connect to System” se

Seite 123 - Environmental Page 113

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 17 for the current session. There are a maximum of eight available COM ports.

Seite 124 - 9-14. Chloride

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 18 3-5. Understanding the Firmware of the ADV6600 There are three sets of

Seite 125 - 9-15. Turbidity

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 19 Select Change System Settings to display the Quick Setup dialog box as sho

Seite 126 - 9-15.1. Effect of Fouling

ADV6600 Y S I Environmental iii Table of Contents Table of Contents ...

Seite 127 - 9-16. Chlorophyll

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 20 appears and then typing Y. To return to the Main menu of the sonde menu,

Seite 128 - 9-16.1. In Vivo Measurement

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 21 3-6.3. Setting Up Installed Sensors To activate the water quality senso

Seite 129 - 9-16.2. Effect of Fouling

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 22 The asterisks (*) that follow the numbers or letters indicat

Seite 130 - 9-16.5. Effect of Turbidity

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 23 calculated parameters after the fact. For example, if you want to log spe

Seite 131 - Environmental Page 121

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 24 DOsat % Local Dissolved oxygen in % air saturation (set to 100 % at c

Seite 132 - 9-17. Rhodamine WT

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 25 Parameter Default Operating Range Comments NO3 A 2.543 Not checked Cl

Seite 133 - Environmental Page 123

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 26 The number of items on this menu depends on the sensors that are availab

Seite 134 - 9-17.2. Effect of Turbidity

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 27 automatically. The wiper functions bi-directionally. A selection of “1”

Seite 135 - 10-1.2. ADV

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 28 mathematically processed to minimize the effect of unusual (or “bad”) read

Seite 136 - Environmental Page 126

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 29 Under most circumstances, the user is advised to make sure that all of t

Seite 137 - 10-1.7. Optical Probes

Table of Contents ADV6600 Y S I Environmental Page iv 3-4. Connecting to the System...

Seite 138 - 10-2.1. O-Rings

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 30 Amplitude The level of returned signal based on the concentration and/or

Seite 139 - 10-2.2. Probe Ports

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 31 Periodically, YSI will make improvements and modifications to the firmwar

Seite 140 - 10-3.3. DO Probe

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 32 11. Choose a Binary update. 12. If you wish to have the firmware update

Seite 141 - Environmental Page 131

Section 3. Installation and Setup ADV6600 Y S I Environmental Page 33 16. Verify the Comm port and Baud settings and make certain that the “Thi

Seite 143 - 10-5. Long-term Storage

ADV6600 Y S I Environmental Page 35 Section 4. Calibration and Diagnostics All of the sonde sensors, with the exception of temperature, r

Seite 144 - 10-5.6. pH and ORP Probes

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 36 4-2.1. Laboratory Beam Check Procedure 1. Connect the ADV6600 to a

Seite 145 - 10-5.8. Optical Probes

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 37 8. Click on the “Record” button so that the data acqu

Seite 146

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 38 4-2.2. Creating a Beam Check File As noted above, the Beam Check fu

Seite 147 - 11-1. Calibration Errors

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 39 A display similar to that shown will be presented. D

Seite 148 - Environmental Page 138

Table of Contents ADV6600 Y S I Environmental Page v 4-6.7. Ammonium ...

Seite 149 - Malfunctioning transmitter

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 40 4-2.4. Viewing a Recorded Beam Check File At the top of the Beam Ch

Seite 150 - Damaged receiver arm

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 41 With these basic facts in mind, it becomes easier t

Seite 151 - 11-3.2. Noisy Velocity Data

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 42 indicates that (a) Doppler pulses are being sent out properly from th

Seite 152 - Environmental Page 142

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 43 To record real-time beam check data for later playback: • Start the

Seite 153

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 44 Compass Calibration Procedure 1. Connect the ADV6600 to a comput

Seite 154

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 45 8. Now cock the sonde at an angle of about 30 degrees toward your bo

Seite 155 - 12-1. Warranty

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 46 NOTE CAREFULLY: Immersion of the pressure sensor to depths of greater

Seite 156 - 12-4. Cleaning Instructions

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 47 4-5.1. Health and Safety Reagents that are used to calibrate and c

Seite 157 - Environmental Page 147

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 48 • Be very cautious that the apparatus does not tip over. • Perfor

Seite 158

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 49 With the ADV6600 in an upright position: 1. Ensure that a gasket i

Seite 159 - Product Return Form

Table of Contents ADV6600 Y S I Environmental Page vi 8-5.1. File...

Seite 160

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 50 6. Select option 1-Calibrate to access calibrations for the water qu

Seite 161 - Environmental Page 151

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 51 4-6.3. Dissolved Oxygen for Unattended Monitoring Studies NOTE CARE

Seite 162

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 52 mg/L Calibration in a Known Solution: 1. Determine a known DO co

Seite 163 - Standards

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 53 A 1-point calibration is used to adjust a previous calibration, gener

Seite 164 - Environmental Page 154

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 54 6. Press Enter again to return to the Calibrate menu. ORP readings

Seite 165 - Environmental Page 155

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 55 A 2-point calibration is performed using two calibration standards th

Seite 166 - A-4. Reagents

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 56 The ammonium sensor should not be exposed to pH calibration buffers.

Seite 167 - Appendix B. Specifications

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 57 equilibration before proceeding. (Allow at least five minutes for t

Seite 168 - Environmental Page 158

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 58 A 2-point calibration is performed using two calibration standards th

Seite 169 - Environmental Page 159

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 59 4-6.10. Turbidity Before calibrating your 6026 or 6136 turbidity s

Seite 170

Table of Contents ADV6600 Y S I Environmental Page vii 9-16.4. Effect of Particle Size ...

Seite 171 - Environmental Page 161

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 60 contaminate the standards during your calibration protocol and cause

Seite 172 - Environmental Page 162

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 61 4-6.11. Chlorophyll To properly calibrate the YSI 6025 Chlorophyll

Seite 173 - Environmental Page 163

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 62 Fax 937-886-9300 www.brightdyes.com 1. Accurately transfer 5.0 g of

Seite 174 - Environmental Page 164

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 63 periodic 2-point calibrations using a dye sample more frequently duri

Seite 175 - Environmental Page 165

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 64 4. Press Enter and the screen will display real-time readings that w

Seite 176 - D-3. Installing the System

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 65 RHODAMINE WT LIQUID (Part # 70301027). As purchased, the solution is

Seite 177 - Environmental Page 167

Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 66 3. Input the value of the first standard at the prompt. No matter h

Seite 178 - D-4. Applications

ADV6600 Y S I Environmental Page 67 Section 5. Field Mounting and Installation Before deploying your ADV6600 in actual field applications

Seite 179 - Environmental Page 169

Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 68 compensate for limited magnetic disturbances, and, if possible, the cal

Seite 180 - Environmental Page 170

Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 69 However, if the deployment requires a special mounting configuration,

Seite 181 - D-6. Software and Firmware

Table of Contents ADV6600 Y S I Environmental Page viii 11-1.2. Out of Range...

Seite 182 - Environmental Page 172

Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 70 5-1.2. Deploying in a PVC Pipe A piece of commonly available 6 inch P

Seite 183 - Environmental Page 173

Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 71 ADV6600 Deployed in Pipe ADV6600 Deployed in Pipe

Seite 184 - D-8. Technical Support

Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 72 5-1.3. Deploying on a Simple Tether The ADV6600 can also be deployed

Seite 185 - Environmental Page 175

Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 73 Some points to keep in mind regarding non-fixed (tethered) deployments

Seite 186

Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 74 This example clearly shows the importance of performin

Seite 187 - Environmental Page 177

Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 75 5-2.2. Flow Interference As noted above, the ADV6600 measures velocit

Seite 189 - Environmental Page 179

ADV6600 Y S I Environmental Page 77 Section 6. Using Your ADV6600 In The Field After setting up your ADV6600, calibrating the sensors, an

Seite 190 - Environmental Page 180

Section 6. Data Collection ADV6600 Y S I Environmental Page 78 6-2. Real-time Data Collection Real-time data collection is an application wher

Seite 191 - Environmental Page 181

Section 6. Data Collection ADV6600 Y S I Environmental Page 79 second warm-up, readings will be logged to the file on the computer hard drive an

Seite 192 - Environmental Page 182

Table of Contents ADV6600 Y S I Environmental Page ix Appendix E. Chlorophyll Measurements...

Seite 193 - Environmental Page 183

Section 6. Data Collection ADV6600 Y S I Environmental Page 80 Before beginning a deployment, make sure you have read and completed the procedur

Seite 194 - Environmental Page 184

Section 6. Data Collection ADV6600 Y S I Environmental Page 81 13. If you chose a.): The data display screen shown in previous section will ap

Seite 195 - F-1. “DOsat %” Convention

Section 6. Data Collection ADV6600 Y S I Environmental Page 82 6-4. SDI-12 Data Collection An SDI-12 data collection application is one in wh

Seite 196 - Environmental Page 186

Section 6. Data Collection ADV6600 Y S I Environmental Page 83 8. Click on Update. The settings will be saved and the Qu

Seite 197 - Environmental Page 187

Section 6. Data Collection ADV6600 Y S I Environmental Page 84 10. Select a unit preference (metric or English) and enter the SDI-12 address t

Seite 198 - Environmental Page 188

Section 6. Data Collection ADV6600 Y S I Environmental Page 85 NOTE: If you have stored your data to the internal recorder during an SDI-12 st

Seite 200 - Environmental Page 190

ADV6600 Y S I Environmental Page 87 Section 7. Downloading ADV6600 Data 7-1. Important Information on Downloading Data Data stored in t

Seite 201

Section 7. Downloading ADV6600 Data ADV6600 Y S I Environmental Page 88 • At the top of the window is a listing of the fi

Seite 202 - Environmental Page 192

Section 7. Downloading ADV6600 Data ADV6600 Y S I Environmental Page 89 • The Destination folder is the location on your computer hard drive o