Operations Manual YSI ADV6600 Environmental Monitoring System
ADV6600 Y S I Environmental Page 91 Section 8. ADVantage 6600 Software – Post-Processing of Data 8-1. Opening a Saved File After a file
Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 92 format of the default template. Note, however, that each time you open a
Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 93 7. The software will now ask you to choose a name for the new file and s
Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 94 1 Left Click and Hold Moves marker to change data point being displayed
Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 95 8-4. Toolbars ADVantage 6600 software utilizes toolbars for some of the
Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 96 8-4.4. Data Collection Toolbar Utilities Future feature – not present
Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 97 Report Opens the Report window for viewing of File Summary, User Settings
Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 98 Tabular Data Displays or hide the Tabular Data at the bottom of the scre
Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 99 Hardware Configuration Displays hardware specifications of the instrumen
ADV6600 Y S I Environmental Page 1 Section 1. Introduction to the ADV6600 1-1. About the ADV6600 The ADV6600 is a fully integrated syste
Section 8. ADVantage 6600 Software ADV6600 Y S I Environmental Page 100 8-5.6. Window ADVantage 6600 software allows for multiple data files t
ADV6600 Y S I Environmental Page 101 Section 9. Principles of Operation 9-1. Acoustic Doppler Velocimeter (ADV) The ADV6600 measures the
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 102 (sediment, small organisms, bubbles, etc.). Some portion of the reflec
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 103 9-2.2. XYZ Using the XYZ coordinate system, velocity measurements are
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 104 Where: dBar = measured pressure (decibar) PressOffset = offset calibra
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 105 mS/cm, which have been prepared in accordance with recommendation 56-19
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 106 9-8. TDS Procedure from Standard Methods for the Examination of Wat
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 107 detectors of DO are highly flow-dependent and therefore require externa
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 108 which is necessary to obtain relatively flow independent measurements.
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 109 9-10. pH Procedure from Standard Methods for the Examination of Water
Section 1. Introduction to the ADV6600 ADV6600 Y S I Environmental Page 2 environmental enforcement personnel, officials of water utilities, an
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 110 intercept at Tm (temperature of measurement). Fortunately, the Nernst
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 111 WARNING: Nitrate sensors can only be used at depths of less than 50 f
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 112 indicates the presence of only small quantities of nitrate, it is unlik
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 113 temperature at least 10°C different from the first two calibration poin
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 114 whichever is larger) refers to sampling applications where only minimal
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 115 and the quality of the water. For all monitoring studies using ion sel
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 116 (2) The absolute values of the readings are significantly closer to th
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 117 particles in the optical path at the instant of measurement. For examp
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 118 The limitations of the in vivo method are outlined below and should be
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 119 The unique YSI chlorophyll system available as an option for use with t
Section 1. Introduction to the ADV6600 ADV6600 Y S I Environmental Page 3 If you have any questions about this product or its application, plea
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 120 From our studies, it appears that entry of a value of 1 to 2 % per degr
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 121 As noted above, the measurement of chlorophyll from in vivo fluorescenc
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 122 provide an accuracy specification for chlorophyll measurement made with
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 123 As for chlorophyll, rhodamine WT fluoresces, that is, when irradiated
Section 9. Principles of Operation ADV6600 Y S I Environmental Page 124 9-17.1. Calibration and Effect of Temperature The sonde software offe
ADV6600 Y S I Environmental Page 125 Section 10. Care, Maintenance, and Storage 10-1. Protection from Biological Fouling The ADV6600 can
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 126 Within the United States, it is recommended to use an anti-foul
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 127 have completed painting the conductivity cell of your probe, it
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 128 10-2. Sonde Care and Maintenance When caring for your ADV6600
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 129 2. Draw the o-ring through the grease while pressing the finge
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 130 10-3. Probe Care and Maintenance 10-3.1. ADV Biological gro
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 131 To resurface the probe: 1. First dry the probe tip completely
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 132 If good pH and/or ORP response is still not restored by the ab
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 133 water under examination. A replacement wiper is supplied with
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 134 10-5.3. Temperature Probe No special precautions are required
Section 10. Care, Maintenance, and Storage ADV6600 Y S I Environmental Page 135 The key to pH probe storage, short or long-term, is to make cer
ADV6600 Y S I Environmental Page 137 Section 11. Troubleshooting This section contains troubleshooting information that will be helpful to
Section 11. Troubleshooting ADV6600 Y S I Environmental Page 138 11-2. Communication Problems 11-2.1. Cannot Communicate With ADV6600 In th
Section 11. Troubleshooting ADV6600 Y S I Environmental Page 139 Low scattering strength If there is insufficient scattering material (e.g.,
ADV6600 Y S I Environmental Page 5 Section 2. Preparing the System for Field Studies Before using your ADV6600 in field studies to correla
Section 11. Troubleshooting ADV6600 Y S I Environmental Page 140 Malfunctioning receiver(s) The signal from each receiver should have about t
Section 11. Troubleshooting ADV6600 Y S I Environmental Page 141 Excessive Noise If the electronics noise level is more than 10 counts above i
Section 11. Troubleshooting ADV6600 Y S I Environmental Page 142 • Be certain the sampling volume of the instrument is sufficiently spaced fro
Section 11. Troubleshooting ADV6600 Y S I Environmental Page 143 SYMPTOM POSSIBLE CAUSE ACTION REFProbe not properly calibrated Follow DO
Section 11. Troubleshooting ADV6600 Y S I Environmental Page 144 SYMPTOM POSSIBLE CAUSE ACTION REFConductivity improperly calibrated. Fol
ADV6600 Y S I Environmental Page 145 Section 12. Warranty and Service Information Thank you for purchasing a YSI ADV6600. This instrument
Section 12. Warranty and Service Information ADV6600 Y S I Environmental Page 146 EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WA
Section 12. Warranty and Service Information ADV6600 Y S I Environmental Page 147 ) If exposure has occurred, the customer must certify that d
ADV6600 Y S I Environmental Page 149 Product Return Form Customer Information Organization _______________________ Contact Name ______________
Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 6 • Hold the probe in a vertical position and apply a few drops of electr
ADV6600 Y S I Environmental Page 150 Product Return Form _________________________________ City, State, Zip _____________________ _____________
ADV6600 Y S I Environmental Page 151 Section 13. Additional Support YSI Environmental offers a wide range of customer assistance and techni
ADV6600 Y S I Environmental Page 153 Appendix A. Accessories and Calibration Standards A-1. Probes and Probe Replacement Parts YSI Model Nu
Appendix A. Accessories and Calibration Standards ADV6600 Y S I Environmental Page 154 A-2. Optional Accessories and Replacement Parts YSI Mo
Appendix A. Accessories and Calibration Standards ADV6600 Y S I Environmental Page 155 YSI Model Number Description 6656 ADVantage 6600 Windo
Appendix A. Accessories and Calibration Standards ADV6600 Y S I Environmental Page 156 A-4. Reagents YSI Model Number Description 3161 Cond
ADV6600 Y S I Environmental Page 157 Appendix B. Specifications Sensor Range Resolution Accuracy Velocity ±0.001 to 6 m/s (±0.003 to 19 ft/s
Appendix B. Specifications ADV6600 Y S I Environmental Page 158 Sensor Range Resolution Accuracy Rhodamine WT 0 to 200 µg/L as true dye; 1,00
ADV6600 Y S I Environmental Page 159 Appendix C. Required Notice The Federal Communications Commission defines this product as a computing de
Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 7 • Remove the old DO membrane and clean the probe tip with water and le
ADV6600 Y S I Environmental Page 161 Appendix D. Frequently Asked Questions D-1. System Description and General Questions Can you give m
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 162 Is it possible to get the ADV6600 in an “EDS” configuration where
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 163 Can I use my existing YSI and/or SonTek field cables when communica
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 164 D-2. What does the system measure? What is the difference betwee
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 165 • The sensor must be totally submerged to successfully measure th
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 166 D-3. Installing the System How should I deploy my ADV6600 for lo
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 167 A few examples of good deployment conditions are: 1. Mounting th
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 168 Can I deploy my ADV6600 by attaching it to a metal pole or mount, o
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 169 2. Spot sampling/profiling of these current and water quality para
Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 8 • Roll the o-ring over the end of the probe, being careful not to touc
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 170 Yes. The standard compass that is present in each ADV6600 will auto
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 171 If I download data to my PC periodically in the field and restart
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 172 A Microsoft Windows operating system is required, and it must be Wi
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 173 No. The released version of ADVantage 6600 has no data merging capa
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 174 No. The ADV sensor does not require calibration. However, if the ar
Appendix D. Frequently Asked Questions ADV6600 Y S I Environmental Page 175 YSI offers a wide range of customer assistance and technical suppor
ADV6600 Y S I Environmental Page 177 Appendix E. Chlorophyll Measurements This section is designed to help the user attain the maximum poss
Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 178 It is important for the user to remember that, because of the limit
Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 179 time of field measurement for the chlorophyll values determined by th
Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 9 2-2.2. Bulkhead Diagram ADV Probe = Non-removable Pressure Sensor
Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 180 Are there precautions that I should take when acquiring samples for
Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 181 analytical procedure has begun. Some users prefer to filter the samp
Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 182 If you consistently observe periods in your monitoring study where th
Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 183 Note the significant improvement in the overall noise level of the
Appendix E. Chlorophyll Measurements ADV6600 Y S I Environmental Page 184 temperature at the site is 10 C, the observed chlorophyll readings wi
ADV6600 Y S I Environmental Page 185 Appendix F. Percent Air Saturation The term “Percent Air Saturation” (abbreviated in many applications
Appendix F. Percent Air Saturation ADV6600 Y S I Environmental Page 186 F-2. “DOsat % Local” Convention The second convention has been used
Appendix F. Percent Air Saturation ADV6600 Y S I Environmental Page 187 F-3. Effects of DO mg/L Calibration When a user performs a mg/L cali
Appendix F. Percent Air Saturation ADV6600 Y S I Environmental Page 188 11. Select Change System Settings. A Quick Setup dialog box will app
ADV6600 Y S I Environmental Page 189 Appendix G. Quick Start Deployment Guide 1-Installing the ADVantage 6600 Software A CDROM containing the
Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 10 2-2.3. O-Ring Lubrication Apply a very thin coat of o-ring lubricant,
Appendix G. Quick Start Deployment Guide ADV6600 Y S I Environmental Page 190 To reduce the amount of space taken up by files on the recorder
ADV6600 Y S I Environmental Page 192 YSI Environmental 1700/1725 Brannum Lane Yellow Springs, OH 45387 USA 937.767.7241 937.767.935
Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 11 Any ISE probe can be installed in either of the two ports labeled “3”
Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 12 2-3.2. Preparing the Cable The cable purchased with your ADV6600 has
Section 2. Preparation of the Sonde ADV6600 Y S I Environmental Page 13 2-3.4. Attaching the Cable Attach your ADV6600 to a computer for ini
ADV6600 Y S I Environmental Page 15 Section 3. Configuring Your ADV6600 – Installation of ADVantage 6600 Software and Sonde Firmware Set-up
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 16 3-4. Connecting to the System When choosing the “Connect to System” se
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 17 for the current session. There are a maximum of eight available COM ports.
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 18 3-5. Understanding the Firmware of the ADV6600 There are three sets of
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 19 Select Change System Settings to display the Quick Setup dialog box as sho
ADV6600 Y S I Environmental iii Table of Contents Table of Contents ...
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 20 appears and then typing Y. To return to the Main menu of the sonde menu,
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 21 3-6.3. Setting Up Installed Sensors To activate the water quality senso
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 22 The asterisks (*) that follow the numbers or letters indicat
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 23 calculated parameters after the fact. For example, if you want to log spe
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 24 DOsat % Local Dissolved oxygen in % air saturation (set to 100 % at c
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 25 Parameter Default Operating Range Comments NO3 A 2.543 Not checked Cl
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 26 The number of items on this menu depends on the sensors that are availab
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 27 automatically. The wiper functions bi-directionally. A selection of “1”
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 28 mathematically processed to minimize the effect of unusual (or “bad”) read
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 29 Under most circumstances, the user is advised to make sure that all of t
Table of Contents ADV6600 Y S I Environmental Page iv 3-4. Connecting to the System...
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 30 Amplitude The level of returned signal based on the concentration and/or
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 31 Periodically, YSI will make improvements and modifications to the firmwar
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 32 11. Choose a Binary update. 12. If you wish to have the firmware update
Section 3. Installation and Setup ADV6600 Y S I Environmental Page 33 16. Verify the Comm port and Baud settings and make certain that the “Thi
ADV6600 Y S I Environmental Page 35 Section 4. Calibration and Diagnostics All of the sonde sensors, with the exception of temperature, r
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 36 4-2.1. Laboratory Beam Check Procedure 1. Connect the ADV6600 to a
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 37 8. Click on the “Record” button so that the data acqu
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 38 4-2.2. Creating a Beam Check File As noted above, the Beam Check fu
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 39 A display similar to that shown will be presented. D
Table of Contents ADV6600 Y S I Environmental Page v 4-6.7. Ammonium ...
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 40 4-2.4. Viewing a Recorded Beam Check File At the top of the Beam Ch
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 41 With these basic facts in mind, it becomes easier t
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 42 indicates that (a) Doppler pulses are being sent out properly from th
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 43 To record real-time beam check data for later playback: • Start the
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 44 Compass Calibration Procedure 1. Connect the ADV6600 to a comput
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 45 8. Now cock the sonde at an angle of about 30 degrees toward your bo
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 46 NOTE CAREFULLY: Immersion of the pressure sensor to depths of greater
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 47 4-5.1. Health and Safety Reagents that are used to calibrate and c
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 48 • Be very cautious that the apparatus does not tip over. • Perfor
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 49 With the ADV6600 in an upright position: 1. Ensure that a gasket i
Table of Contents ADV6600 Y S I Environmental Page vi 8-5.1. File...
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 50 6. Select option 1-Calibrate to access calibrations for the water qu
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 51 4-6.3. Dissolved Oxygen for Unattended Monitoring Studies NOTE CARE
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 52 mg/L Calibration in a Known Solution: 1. Determine a known DO co
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 53 A 1-point calibration is used to adjust a previous calibration, gener
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 54 6. Press Enter again to return to the Calibrate menu. ORP readings
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 55 A 2-point calibration is performed using two calibration standards th
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 56 The ammonium sensor should not be exposed to pH calibration buffers.
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 57 equilibration before proceeding. (Allow at least five minutes for t
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 58 A 2-point calibration is performed using two calibration standards th
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 59 4-6.10. Turbidity Before calibrating your 6026 or 6136 turbidity s
Table of Contents ADV6600 Y S I Environmental Page vii 9-16.4. Effect of Particle Size ...
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 60 contaminate the standards during your calibration protocol and cause
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 61 4-6.11. Chlorophyll To properly calibrate the YSI 6025 Chlorophyll
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 62 Fax 937-886-9300 www.brightdyes.com 1. Accurately transfer 5.0 g of
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 63 periodic 2-point calibrations using a dye sample more frequently duri
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 64 4. Press Enter and the screen will display real-time readings that w
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 65 RHODAMINE WT LIQUID (Part # 70301027). As purchased, the solution is
Section 4. Calibration and Diagnostics ADV6600 Y S I Environmental Page 66 3. Input the value of the first standard at the prompt. No matter h
ADV6600 Y S I Environmental Page 67 Section 5. Field Mounting and Installation Before deploying your ADV6600 in actual field applications
Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 68 compensate for limited magnetic disturbances, and, if possible, the cal
Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 69 However, if the deployment requires a special mounting configuration,
Table of Contents ADV6600 Y S I Environmental Page viii 11-1.2. Out of Range...
Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 70 5-1.2. Deploying in a PVC Pipe A piece of commonly available 6 inch P
Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 71 ADV6600 Deployed in Pipe ADV6600 Deployed in Pipe
Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 72 5-1.3. Deploying on a Simple Tether The ADV6600 can also be deployed
Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 73 Some points to keep in mind regarding non-fixed (tethered) deployments
Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 74 This example clearly shows the importance of performin
Section 5. Mounting and Installation ADV6600 Y S I Environmental Page 75 5-2.2. Flow Interference As noted above, the ADV6600 measures velocit
ADV6600 Y S I Environmental Page 77 Section 6. Using Your ADV6600 In The Field After setting up your ADV6600, calibrating the sensors, an
Section 6. Data Collection ADV6600 Y S I Environmental Page 78 6-2. Real-time Data Collection Real-time data collection is an application wher
Section 6. Data Collection ADV6600 Y S I Environmental Page 79 second warm-up, readings will be logged to the file on the computer hard drive an
Table of Contents ADV6600 Y S I Environmental Page ix Appendix E. Chlorophyll Measurements...
Section 6. Data Collection ADV6600 Y S I Environmental Page 80 Before beginning a deployment, make sure you have read and completed the procedur
Section 6. Data Collection ADV6600 Y S I Environmental Page 81 13. If you chose a.): The data display screen shown in previous section will ap
Section 6. Data Collection ADV6600 Y S I Environmental Page 82 6-4. SDI-12 Data Collection An SDI-12 data collection application is one in wh
Section 6. Data Collection ADV6600 Y S I Environmental Page 83 8. Click on Update. The settings will be saved and the Qu
Section 6. Data Collection ADV6600 Y S I Environmental Page 84 10. Select a unit preference (metric or English) and enter the SDI-12 address t
Section 6. Data Collection ADV6600 Y S I Environmental Page 85 NOTE: If you have stored your data to the internal recorder during an SDI-12 st
ADV6600 Y S I Environmental Page 87 Section 7. Downloading ADV6600 Data 7-1. Important Information on Downloading Data Data stored in t
Section 7. Downloading ADV6600 Data ADV6600 Y S I Environmental Page 88 • At the top of the window is a listing of the fi
Section 7. Downloading ADV6600 Data ADV6600 Y S I Environmental Page 89 • The Destination folder is the location on your computer hard drive o