ba76013e01 01/2012IQ SENSOR NETTriOxmatic® 700 IQ (SW)Operating manualIQ SENSOR NET D.O. sensorTriOxmatic 700 IQTriOxmatic 700 IQ SW

Seite 2 - © 2012 Xylem Inc

Safety TriOxmatic® 700 IQ (SW)2 - 4ba76013e01 01/2012

Seite 3

TriOxmatic® 700 IQ (SW) Commissioning3 - 1ba76013e01 01/20123 Commissioning 3.1 Scope of delivery TriOxmatic® 700 IQ (SW) ZBK 600 accessory kit The

Seite 4 - List of contents TriOxmatic

Commissioning TriOxmatic® 700 IQ (SW)3 - 2ba76013e01 01/2012Connecting the sensor to the sensor connection cableFig. 3-1 Connecting the sensor1 Take t

Seite 5 - 1Overview

TriOxmatic® 700 IQ (SW) Commissioning3 - 3ba76013e01 01/20123.3 Commissioning / Getting the instrument ready for measuring NoteFor the sensor to po

Seite 6

Commissioning TriOxmatic® 700 IQ (SW)3 - 4ba76013e01 01/20123.4 TriOxmatic® 700 IQ (SW) setting tableSetting Selection/values ExplanationMeasuring mod

Seite 7 - 2 Safety

TriOxmatic® 700 IQ (SW) Commissioning3 - 5ba76013e01 01/2012Carrying out settings Using s, switch from the measured value display to the main menu of

Seite 8 - 2.1 Authorized use

Commissioning TriOxmatic® 700 IQ (SW)3 - 6ba76013e01 01/2012

Seite 9 - 700 IQ (SW) Safety

TriOxmatic® 700 IQ (SW) Measuring / Operation4 - 1ba76013e01 01/20124 Measuring / Operation4.1 MeasuringWarningContact with the sample can lead to dan

Seite 10 - Safety TriOxmatic

Measuring / Operation TriOxmatic® 700 IQ (SW)4 - 2ba76013e01 01/2012General course of a calibration on the IQ SENSOR NETGenerally, calibration on the

Seite 11 - 3 Commissioning

TriOxmatic® 700 IQ (SW) Measuring / Operation4 - 3ba76013e01 01/2012NoteUp to this point, you can break off the calibration procedure at any time with

Seite 12 - Commissioning TriOxmatic

TriOxmatic® 700 IQ (SW)2ba76013e01 01/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI 1725 Brannum Lane Yellow

Seite 13 - 700 IQ (SW) Commissioning

Measuring / Operation TriOxmatic® 700 IQ (SW)4 - 4ba76013e01 01/20124.2.3 Calibration historyCalibration history (available in the IQ SENSOR NET syste

Seite 14 - 700 IQ (SW) setting table

TriOxmatic® 700 IQ (SW) Measuring / Operation4 - 5ba76013e01 01/20124.3 Function checkThe function check is the simplest way of letting you know wheth

Seite 15

Measuring / Operation TriOxmatic® 700 IQ (SW)4 - 6ba76013e01 01/2012Measured values within the required precisionIf the measured value lies within the

Seite 16

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 1ba76013e01 01/20125 Maintenance, cleaning, disposal, and replacement5.1 G

Seite 17 - 4 Measuring / Operation

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 2ba76013e01 01/20125.2 Cleaning the sensor shaft and membrane For normal o

Seite 18

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 3ba76013e01 01/2012CautionClean the membrane very carefully! Make sure the

Seite 19

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 4ba76013e01 01/20128 If necessary, carefully dry the membrane with a lint-

Seite 20

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 5ba76013e01 01/20125.3 Changing the electrolyte and membrane capYSI delive

Seite 21 - 4.3 Function check

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 6ba76013e01 01/20124 Unscrew the membrane cap (for disposal of membrane c

Seite 22

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 7ba76013e01 01/20127 Thoroughly rinse the sensor head with electrolyte sol

Seite 23

TriOxmatic® 700 IQ (SW) List of contents0 - 1ba76013e01 01/2012TriOxmatic® 700 IQ (SW) - List of contents1 Overview . . . . . . . . . . . . . . . . .

Seite 24

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 8ba76013e01 01/2012CautionThe ELY/A electrolyte solution irritates eyes, s

Seite 25

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 9ba76013e01 01/2012Precision measurementsRecommendation: In order to carry

Seite 26

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 10ba76013e01 01/20125.4 Cleaning the electrodesThe counter electrode and r

Seite 27

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 11ba76013e01 01/2012CautionThe ELY/A electrolyte solution irritates eyes,

Seite 28

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 12ba76013e01 01/20125.4.1 Cleaning the gold working electrodeCautionDo not

Seite 29

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 13ba76013e01 01/20125.4.2 Cleaning the silver counter electrodeWe recommen

Seite 30

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 14ba76013e01 01/20123 Remove the screw cap from the safety cap of the clea

Seite 31

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 15ba76013e01 01/20127 Unscrew the screw cap.8 Thoroughly rinse the electro

Seite 32 - 5.4 Cleaning the electrodes

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 16ba76013e01 01/201210 Rinse the sensor head and electrode unit several ti

Seite 33

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 17ba76013e01 01/2012Cleaning the silver counter electrode in a beakerCauti

Seite 34

List of contents TriOxmatic® 700 IQ (SW)0 - 2ba76013e01 01/20127 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17.1

Seite 35

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 18ba76013e01 01/20123 Clamp the sensor in a stand. Submerse the electrode

Seite 36

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 19ba76013e01 01/20127 Carefully shake off the drops of water.8 Fill a new

Seite 37 - 9 Unscrew the safety cap

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 20ba76013e01 01/20125.5 Checking the sensor for freedom from zero-current

Seite 38

TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 21ba76013e01 01/20125.6 StorageStore the clean and dry sensor with its pro

Seite 39

Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 22ba76013e01 01/20125.8 Maintenance equipment and replacement partsNoteInf

Seite 40

TriOxmatic® 700 IQ (SW) What to do if...6 - 1ba76013e01 01/20126 What to do if...The sensor is in the air and the display shows 0.0 mg/l or 0 % O2The

Seite 41

What to do if... TriOxmatic® 700 IQ (SW)6 - 2ba76013e01 01/2012Measured value fluctuating heavilyMeasured values too highIncorrect temperature display

Seite 42

TriOxmatic® 700 IQ (SW) Technical data7 - 1ba76013e01 01/20127 Technical data7.1 Measurement characteristicsMeasuring principle Membrane-covered amper

Seite 43 - 5.7 Disposal

Technical data TriOxmatic® 700 IQ (SW)7 - 2ba76013e01 01/2012Air pressure compensation7.2 Application characteristicsAllowed temperature rangeAllowed

Seite 44

TriOxmatic® 700 IQ (SW) Technical data7 - 3ba76013e01 01/20127.3 General dataDimensionsWeight (without sensor connection cable)Connection technique Co

Seite 45 - 6 What to do if

TriOxmatic® 700 IQ (SW) Overview1 - 1ba76013e01 01/20121Overview1.1 How to use this component operating manualStructure of the IQ SENSOR NEToperating

Seite 46 - What to do if... TriOxmatic

Technical data TriOxmatic® 700 IQ (SW)7 - 4ba76013e01 01/2012Automatic sensor monitoring (SensCheck function)* Note: The leak recognition can be selec

Seite 47 - 7 Technical data

TriOxmatic® 700 IQ (SW) Contact Information8 - 1ba76013e01 01/20128 Contact Information8.1 Ordering & Technical SupportWhen placing an order pleas

Seite 48 - Technical data TriOxmatic

Contact Information TriOxmatic® 700 IQ (SW)8 - 2ba76013e01 01/2012

Seite 49 - 7.3 General data

TriOxmatic® 700 IQ (SW) Indexes9 - 1ba76013e01 01/20129 Indexes9.1 Explanation of the messagesThis chapter contains a list of all the message codes an

Seite 50 - 7.4 Electrical data

Indexes TriOxmatic® 700 IQ (SW)9 - 2ba76013e01 01/20129.1.2 Info messagesEI1331 Operational voltage too low* Check installation and cable lengths, Fol

Seite 51 - 8 Contact Information

TriOxmatic® 700 IQ (SW) Indexes9 - 3ba76013e01 01/20129.2 Status infoThe status info is a coded piece of information on the current status of a sensor

Seite 52

Indexes TriOxmatic® 700 IQ (SW)9 - 4ba76013e01 01/2012

Seite 54 - Indexes TriOxmatic

1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com

Seite 55 - 9.2 Status info

Overview TriOxmatic® 700 IQ (SW)1 - 2ba76013e01 01/20121.2 Structure of the TriOxmatic® 700 IQ (SW)Fig. 1-2 Structure of the D. O. sensor (example: Tr

Seite 56

TriOxmatic® 700 IQ (SW) Safety2 - 1ba76013e01 01/20122 SafetyThis component operating manual contains special instructions that must be followed durin

Seite 57

Safety TriOxmatic® 700 IQ (SW)2 - 2ba76013e01 01/20122.1 Authorized useThe authorized use of the TriOxmatic® 700 IQ (SW) comprises its use as a D.O. s

Seite 58

TriOxmatic® 700 IQ (SW) Safety2 - 3ba76013e01 01/2012Safe operation If safe operation is no longer possible, the sensor must be taken out of operation