ba76013e01 01/2012IQ SENSOR NETTriOxmatic® 700 IQ (SW)Operating manualIQ SENSOR NET D.O. sensorTriOxmatic 700 IQTriOxmatic 700 IQ SW
Safety TriOxmatic® 700 IQ (SW)2 - 4ba76013e01 01/2012
TriOxmatic® 700 IQ (SW) Commissioning3 - 1ba76013e01 01/20123 Commissioning 3.1 Scope of delivery TriOxmatic® 700 IQ (SW) ZBK 600 accessory kit The
Commissioning TriOxmatic® 700 IQ (SW)3 - 2ba76013e01 01/2012Connecting the sensor to the sensor connection cableFig. 3-1 Connecting the sensor1 Take t
TriOxmatic® 700 IQ (SW) Commissioning3 - 3ba76013e01 01/20123.3 Commissioning / Getting the instrument ready for measuring NoteFor the sensor to po
Commissioning TriOxmatic® 700 IQ (SW)3 - 4ba76013e01 01/20123.4 TriOxmatic® 700 IQ (SW) setting tableSetting Selection/values ExplanationMeasuring mod
TriOxmatic® 700 IQ (SW) Commissioning3 - 5ba76013e01 01/2012Carrying out settings Using s, switch from the measured value display to the main menu of
Commissioning TriOxmatic® 700 IQ (SW)3 - 6ba76013e01 01/2012
TriOxmatic® 700 IQ (SW) Measuring / Operation4 - 1ba76013e01 01/20124 Measuring / Operation4.1 MeasuringWarningContact with the sample can lead to dan
Measuring / Operation TriOxmatic® 700 IQ (SW)4 - 2ba76013e01 01/2012General course of a calibration on the IQ SENSOR NETGenerally, calibration on the
TriOxmatic® 700 IQ (SW) Measuring / Operation4 - 3ba76013e01 01/2012NoteUp to this point, you can break off the calibration procedure at any time with
TriOxmatic® 700 IQ (SW)2ba76013e01 01/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI 1725 Brannum Lane Yellow
Measuring / Operation TriOxmatic® 700 IQ (SW)4 - 4ba76013e01 01/20124.2.3 Calibration historyCalibration history (available in the IQ SENSOR NET syste
TriOxmatic® 700 IQ (SW) Measuring / Operation4 - 5ba76013e01 01/20124.3 Function checkThe function check is the simplest way of letting you know wheth
Measuring / Operation TriOxmatic® 700 IQ (SW)4 - 6ba76013e01 01/2012Measured values within the required precisionIf the measured value lies within the
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 1ba76013e01 01/20125 Maintenance, cleaning, disposal, and replacement5.1 G
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 2ba76013e01 01/20125.2 Cleaning the sensor shaft and membrane For normal o
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 3ba76013e01 01/2012CautionClean the membrane very carefully! Make sure the
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 4ba76013e01 01/20128 If necessary, carefully dry the membrane with a lint-
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 5ba76013e01 01/20125.3 Changing the electrolyte and membrane capYSI delive
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 6ba76013e01 01/20124 Unscrew the membrane cap (for disposal of membrane c
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 7ba76013e01 01/20127 Thoroughly rinse the sensor head with electrolyte sol
TriOxmatic® 700 IQ (SW) List of contents0 - 1ba76013e01 01/2012TriOxmatic® 700 IQ (SW) - List of contents1 Overview . . . . . . . . . . . . . . . . .
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 8ba76013e01 01/2012CautionThe ELY/A electrolyte solution irritates eyes, s
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 9ba76013e01 01/2012Precision measurementsRecommendation: In order to carry
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 10ba76013e01 01/20125.4 Cleaning the electrodesThe counter electrode and r
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 11ba76013e01 01/2012CautionThe ELY/A electrolyte solution irritates eyes,
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 12ba76013e01 01/20125.4.1 Cleaning the gold working electrodeCautionDo not
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 13ba76013e01 01/20125.4.2 Cleaning the silver counter electrodeWe recommen
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 14ba76013e01 01/20123 Remove the screw cap from the safety cap of the clea
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 15ba76013e01 01/20127 Unscrew the screw cap.8 Thoroughly rinse the electro
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 16ba76013e01 01/201210 Rinse the sensor head and electrode unit several ti
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 17ba76013e01 01/2012Cleaning the silver counter electrode in a beakerCauti
List of contents TriOxmatic® 700 IQ (SW)0 - 2ba76013e01 01/20127 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17.1
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 18ba76013e01 01/20123 Clamp the sensor in a stand. Submerse the electrode
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 19ba76013e01 01/20127 Carefully shake off the drops of water.8 Fill a new
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 20ba76013e01 01/20125.5 Checking the sensor for freedom from zero-current
TriOxmatic® 700 IQ (SW) Maintenance, cleaning, disposal, and replacement5 - 21ba76013e01 01/20125.6 StorageStore the clean and dry sensor with its pro
Maintenance, cleaning, disposal, and replacement TriOxmatic® 700 IQ (SW)5 - 22ba76013e01 01/20125.8 Maintenance equipment and replacement partsNoteInf
TriOxmatic® 700 IQ (SW) What to do if...6 - 1ba76013e01 01/20126 What to do if...The sensor is in the air and the display shows 0.0 mg/l or 0 % O2The
What to do if... TriOxmatic® 700 IQ (SW)6 - 2ba76013e01 01/2012Measured value fluctuating heavilyMeasured values too highIncorrect temperature display
TriOxmatic® 700 IQ (SW) Technical data7 - 1ba76013e01 01/20127 Technical data7.1 Measurement characteristicsMeasuring principle Membrane-covered amper
Technical data TriOxmatic® 700 IQ (SW)7 - 2ba76013e01 01/2012Air pressure compensation7.2 Application characteristicsAllowed temperature rangeAllowed
TriOxmatic® 700 IQ (SW) Technical data7 - 3ba76013e01 01/20127.3 General dataDimensionsWeight (without sensor connection cable)Connection technique Co
TriOxmatic® 700 IQ (SW) Overview1 - 1ba76013e01 01/20121Overview1.1 How to use this component operating manualStructure of the IQ SENSOR NEToperating
Technical data TriOxmatic® 700 IQ (SW)7 - 4ba76013e01 01/2012Automatic sensor monitoring (SensCheck function)* Note: The leak recognition can be selec
TriOxmatic® 700 IQ (SW) Contact Information8 - 1ba76013e01 01/20128 Contact Information8.1 Ordering & Technical SupportWhen placing an order pleas
Contact Information TriOxmatic® 700 IQ (SW)8 - 2ba76013e01 01/2012
TriOxmatic® 700 IQ (SW) Indexes9 - 1ba76013e01 01/20129 Indexes9.1 Explanation of the messagesThis chapter contains a list of all the message codes an
Indexes TriOxmatic® 700 IQ (SW)9 - 2ba76013e01 01/20129.1.2 Info messagesEI1331 Operational voltage too low* Check installation and cable lengths, Fol
TriOxmatic® 700 IQ (SW) Indexes9 - 3ba76013e01 01/20129.2 Status infoThe status info is a coded piece of information on the current status of a sensor
Indexes TriOxmatic® 700 IQ (SW)9 - 4ba76013e01 01/2012
1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com
Overview TriOxmatic® 700 IQ (SW)1 - 2ba76013e01 01/20121.2 Structure of the TriOxmatic® 700 IQ (SW)Fig. 1-2 Structure of the D. O. sensor (example: Tr
TriOxmatic® 700 IQ (SW) Safety2 - 1ba76013e01 01/20122 SafetyThis component operating manual contains special instructions that must be followed durin
Safety TriOxmatic® 700 IQ (SW)2 - 2ba76013e01 01/20122.1 Authorized useThe authorized use of the TriOxmatic® 700 IQ (SW) comprises its use as a D.O. s
TriOxmatic® 700 IQ (SW) Safety2 - 3ba76013e01 01/2012Safe operation If safe operation is no longer possible, the sensor must be taken out of operation